Proper controls are essential for assessing staining specificity. Negative controls ensure that the observed signals are due to the presence of the target antigen and not due to non-specific binding or other artifacts. SYSY Antibodies recommends the following control experiments to ensure staining specificity.
In chromogenic IHC, endogenous peroxidases or alkaline phosphatases can cause false positive staining if not properly blocked. Omitting both the primary antibody and the secondary system in the staining protocol reveals endogenous enzyme activity.
Certain tissue-inherent pigments may also be mistaken for DAB signals due to their yellow-brown color, including:
Figure 1: Siderin (arrows, yellow) in formalin fixed paraffin embedded (FFPE) mouse spleen. Nuclei have been visualized by hematoxylin staining (blue).
Figure 2: Melanin (arrow, brown) in formalin fixed paraffin embedded (FFPE) mouse retina. Nuclei have been visualized by hematoxylin staining (blue).
In IF, autofluorescence is a widely occurring phenomenon in cells and tissues caused by endogenous molecules such as:
Figure 3: Autofluorescence in formalin fixed paraffin embedded (FFPE) human placenta. Autofluorescence of red blood cells is shown in green. Nuclei have been visualized by DAPI staining (blue).
Figure 4: Autofluorescence in formalin fixed paraffin embedded (FFPE) human spleen. Autofluorescence of red blood cells is shown in red. Nuclei have been visualized by DAPI staining (blue).
Naturally fluorescent components in tissues can emit signals that overlap with fluorophore spectra, reducing the signal-to-noise ratio and complicating image interpretation. For example, collagen and NADH primarily emit in the blue to green range. Therefore, selecting fluorophores with emission in the red or far-red spectrum can help minimize spectral overlap. Lipofuscin, however, fluoresces broadly across the spectrum—most strongly between 500 and 695 nm—and can be particularly problematic, as its granular appearance may be mistaken for specific staining.
Autofluorescence can also arise from aldehyde-based fixation methods, such as formaldehyde or glutaraldehyde. Formalin-induced autofluorescence produces broad emission across a wide spectral range, including the blue, green, and red regions, further complicating signal detection. Consequently, appropriate controls and autofluorescence-reduction strategies are essential to ensure accurate analysis.
Negative controls omitting the primary antibody help to identify false positive staining caused by the secondary system.
Suitable samples include:
Figure 8: Validation of species-specific antibodies using tissues from positive and negative species. Indirect immunostaining of formalin fixed paraffin embedded (FFPE) sections of human placenta (A) and mouse placenta (B) with rabbit anti-human Lamin B1 antibody (cat. no. HS-404 018, dilution 1:1000, DAB, brown). Nuclei have been visualized by hematoxylin staining (blue).
This antibody is suited to distinguish human cells from murine cells in humanized mouse models.
Figure 9: Validation of primary antibodies using knockdown models. Indirect immunostaining of formalin fixed paraffin embedded (FFPE) mouse lung sections of wild type (A) or LAMP3 knock-out (B) mice with rat anti-LAMP3 antibody (cat. no. HS-391 017, dilution 1:000, DAB, brown). Nuclei have been visualized by hematoxylin staining (blue).
Courtesy: Knock-out mouse lung tissue was kindly provided by Prof. Dr. Markus Damme, Christian-Albrechts-University of Kiel, Kiel, Germany